Centre for Clinical Trials and Research (CCTR)
Digital PCR QIAcuity One 5 Plex
This nanoplate-based digital PCR system gives absolute quantification of DNA and RNA targets without a standard curve. It detects up to five targets in a single run using five-colour multiplexing. Each well is divided into about 8,500 or 26,000 partitions (8.5k or 26k nanoplates), and the whole workflow of partitioning, thermocycling and imaging takes about 2 hours. It suits rare mutation detection, copy number variation, viral load and gene expression studies.
PCR – Thermofisher Scientific – ProflexTM Base
This thermal cycler has three independently controlled blocks, so three different reactions can be run at the same time. It runs from 0 to 100 °C with ±0.25 °C accuracy and block uniformity better than 0.5 °C. It is used for DNA/RNA amplification, genotyping, cloning and sequencing preparation, and it lets users run different protocols or optimise annealing temperatures in a single session.
Agarose Gel Electrophoresis - BIORAD
AGE separates DNA and RNA fragments by size, typically from about 100 bp to over 10 kb depending on the gel percentage (0.7–2%). Users can check PCR products, fragment size and nucleic acid integrity before further analysis. It is a simple, quick and cost-effective quality check.
NanoDrop Lite Plus
AGE separates DNA and RNA fragments by size, typically from about 100 bp to over 10 kb depending on the gel percentage (0.7–2%). Users can check PCR products, fragment size and nucleic acid integrity before further analysis. It is a simple, quick and cost-effective quality check.
Western Blotting BIORAD
This system separates proteins by size using SDS-PAGE and transfers them to a membrane for antibody-based detection. It typically resolves proteins in the range of about 10–250 kDa. It confirms protein identity, relative expression and post-translational modifications, and is widely used for biomarker validation and disease studies.
Chemi Doc Imaging system
The ChemiDoc captures images of gels and blots using chemiluminescent, fluorescent and colorimetric detection. Its imaging software allows band quantification and molecular weight estimation. It gives sensitive, reproducible and publication-quality documentation of results.
ELISA reader BIORAD
This microplate absorbance reader works with 96-well plates and measures between 400 and 750 nm. It comes with 415, 450, 490, 595, 655 and 750 nm filters, and a plate is read in about 6 seconds at a single wavelength. It is used for ELISA, protein assays and kinetic assays to quantify hormones, antibodies, cytokines and biomarkers.
Hormone Analyser – Accre Peerless
This is a bench-top chemiluminescence immunoassay (CLIA) analyser. It measures thyroid markers (TSH, T3, T4, fT3, fT4) and reproductive hormones (LH, FSH, estradiol, prolactin, testosterone, progesterone, β-hCG). The first result is typically ready in about 10 minutes, and the automated processing reduces manual error.
Autoanalyser – ErbaXL640 – Transasia Bio-Medical Ltd
This fully automated, random-access clinical chemistry analyser runs up to 400 photometric tests per hour, or 640 tests per hour with the optional ISE. It has a 12-wavelength photometer (340–750 nm) with an absorbance range of 0–3.0 OD. With the ISE module, it tests glucose, lipid profile, liver and kidney function markers, and electrolytes (Na, K, Cl, Li).
Hematology Analyser – Horiba Medical – Yumizen H500 OT
This analyser gives a complete blood count with 5-part differential, covering RBC, WBC, platelets, haemoglobin and related indices. It runs up to 60 samples per hour using only about 20 µL of blood in open-tube mode. It is used to assess health status, detect haematological abnormalities and monitor treatment effects.
pH Meter – Mettler Toledo – Five Easy Plus FP20
This benchtop meter measures pH from -2.00 to 16.00 with ±0.01 accuracy and 0.01/0.1 resolution. It also measures mV (±2000 mV) and temperature (-5 to 105 °C), with automatic or manual temperature compensation. It gives accurate pH for buffers, reagents and biological samples.